These models allow for the user to calibrate the IRt/c’s to match a thermocouple in whatever temperature range they desire. The range adjustment screw is adjusted to calibrate for temperature range of interest and material properties (emissivity). Models for non-metals (HiE) and metals (LoE) are available.
ORDERING INFORMATION : (Model)-(Thermocouple Type) - (HE or LoE) Select the model for temperature range , sport size , thermocouple type desired (J.K.R or S) and target
Example IRt/c 10A-K-HiE
ATTENTION OEMS - All of the adjustable model available precalibrated from Exergen with NIST traceability for quick multiple same sensor installations. Also the millivolt output curves, mv table, and Eurothern downloadable files are available for all sensor models for special calibration
FULL MODEL SPECIFICATIONS
Also see related Technotes 58 ,59 ,60 ,61 ,66 ,69 ,70 ,73 ,76
Spot size = 2* D* Tan ø + Min Spot Size
D = T-F
T = Total Distance between sensor and target
F = Distance of Focal Point
These models allow for the user to calibrate the IRt/c’s to match a thermocouple in the temperature range they desire. The range adjustment screw is adjusted to calibrate for temperature range of interest and material properties (emissivity). Models for non-metals (HiE) and metals (LoE) are available.
GENERAL SPECIFICATIONS
same as precalibrated models, with the exception of emissivity, which is 0.2 for LoE models and 0.9 for HiE models
ORDERING INFORMATION
(Model)-(Thermocouple Type) - (HiE or LoE)
Select the model for temperature range , spot size , thermocouple type desired (J.K or.R/ S) and target
Contact 617.923.9900, ext 6238 or 6263, email: industrial@exergen.com and fill out our TECH SUPPORT Formand an expert will be in touch with you within 1 hour (during factory hours of 8AM – 5PM, US EST)