| IRt/c Principles of Operation - The IRt/c is calibrated to match a conventional thermocouple at a specific temperature. By adjusting the calibration setting (c) of the IRt/c to equal the Seebeck coefficient (α), the signal output (mVout) will be within 2% linear to match the conventional t/c at temperatures close to the calibration point. So when the non-contact, self powered IRt/c is wired into a conventional t/c input device the target temperature (Tt) displayed will be the same as the surface temperature (Ts), the conventional t/c is displaying. |
| Target controlled to 240F) |
|
|
| Conventional t/c
mVout = a (Ts-Tcj)
a = Seebeck coefficient |
| IRt/c
mVout = c (Tt-Ts)+a (Ts-Tcj)
when c=a
mVout = (Tt-Tcj) c = calibration setting |
|
|
|
|
|
|
|
|
|
|