| #10 IRt/c's are Intrinsically Safe When Used with Barriers |
| "Field Apparatus having energy storing or generating characteristics of 1.2V, 0.1A, 25 mW or 25 microJ shall be considered Simple Apparatus (non-energy storing). These general purpose devices may be used in a hazardous (classified) location without further approval when connected to a certified intrinsically safe circuit." - Quote from R. Stahl, Inc. Comprehensive Product Manual On Intrinsic Safety Barrier and Repeater Relays.Examples of non-energy storing Intrinsically Safe Apparatus are: |
|
 |
Thermocouples, RTD's, LED's |
 |
Dry Switch Contacts |
 |
NAMUR Inductive Proximity Switches |
 |
Non-inductive Strain Gauge Devices and Resistors |
|
| The IRt/c falls into the category of thermocouples, since it generates its signal by converting the radiated heat energy to an electrical signal via Seebeck effects, the basic driving force of thermo-couples. Like all thermocouples, it requires no power source and generates signals measured in millivolts of voltage, microamps of current and nanowatts of power. IRt/c's have a small capacitance, but at one microFarad, the energy storage is measured in nanojoules and is a thousand times lower than the 25 microjoule criterion. Accordingly, the IRt/c qualifies as a Simple Apparatus for use in hazardous locations, and with the appropriate barrier, qualifies as Intrinsically Safe. |
|
|
|